How is an ISO 8 cleanroom classified?
The 8 in “ISO 8 cleanroom” designates an air cleanliness class by particle concentration under ISO 14644-1:2015. It measures one thing only: the concentration of airborne particles in the room.
ISO 14644-1 classifies the air cleanliness of cleanrooms and clean zones by this concentration, for particles whose size threshold lies between 0.1 µm and 5 µm. Its reference method is the light-scattering airborne particle counter. The 2nd edition dates from December 2015. ISO confirmed it in 2021. When we checked on 23 November 2026, it was under systematic review.
ISO also states what the standard does not do. It does not characterise the physical, chemical, radiological or viable nature of the particles. On its own, an ISO 8 classification therefore says nothing about the micro-organisms present.
“Cleanroom” and “clean room” mean the same thing. The standard uses the second term.
3,520,000 particles per cubic metre at rest
At rest, a Grade D room allows at most 3,520,000 particles of 0.5 µm and larger per cubic metre of air. For particles of 5 µm and larger the limit is 29,300.
These values come from Table 1 of Annex 1 to European good manufacturing practice (GMP), in its 2022 version. In operation, the annex sets no limit for Grade D: the manufacturer establishes one from a risk analysis and its routine data. The limits specific to the ISO 8 class are in ISO 14644-1, which is not freely available.
Table 1 of Annex 1 shows where Grade D stands among the 4 GMP grades (maximum number of particles per cubic metre of air):
| Grade | 0.5 µm and larger, at rest | 0.5 µm and larger, in operation | 5 µm and larger, at rest |
|---|---|---|---|
| A | 3,520 | 3,520 | not specified |
| B | 3,520 | 352,000 | not specified |
| C | 352,000 | 3,520,000 | 2,930 |
| D | 3,520,000 | set by the manufacturer | 29,300 |
The annex defines the 2 measurement states. “At rest”: installations running, equipment in place but not operating, no personnel. “In operation”: equipment running, with the maximum number of people planned. After operations, the at-rest limits must be recovered after a short clean-up period, whose guidance value is less than 20 minutes.
Where does Grade D come from?
Grades A, B, C and D belong to Annex 1 of European GMP, on the manufacture of sterile medicinal products. At rest, Grade D is comparable to ISO class 8, hence the double name “ISO 8 (Grade D)”. This correspondence appears in the good preparation practices (2023) of France’s National Agency for Medicines and Health Products Safety (ANSM), which likewise compares Grades A and B to ISO class 5.
Annex 1 is part of Volume 4 of EudraLex. The European Commission’s revision, dated 22 August 2022, has applied since 25 August 2023. In France, ANSM took it up in its GMP guide by a decision of 28 May 2024, in force since 14 June 2024.
The legal basis of Annex 1 is medicines law (Article 47 of Directive 2001/83/EC). For a medical device it is not the applicable text. It remains the source of Grade D. Its scope allows some of its principles to be used for other products not intended to be sterile: cleanroom classification, qualification, monitoring and personnel gowning, among others. The manufacturer then documents the principles it applies.
Grade D suits the less critical steps, not aseptic filling
Annex 1 describes Grades C and D as clean areas for less critical stages in the manufacture of aseptically filled sterile products. Its examples of Grade D operations: cleaning of equipment, handling of components after washing, and preparation of solutions and accessories before filling.
Our ISO 8 (Grade D) cleanroom is dedicated to packaging. The workshop also includes:
- a zone dedicated to secondary packaging
- a preparation zone
- a manufacturing zone with controlled temperature and humidity
- separate storage for work in progress and finished products
Only one batch at a time is handled at each workstation. We clean parts following your protocols, then carry out primary and secondary packaging, labelling, and the creation and insertion of instructions for use, with each batch followed in a record that runs from raw material to shipment. The detail of the operations is on our page on ISO 8 cleanroom packaging.
Does an ISO 8 room make a device sterile?
No. An ISO 8 room limits airborne particles: it does not replace any sterilisation step.
In GMP, high-risk aseptic operations such as filling are carried out in Grade A, within a Grade B background. Grade D is not used for those operations.
If your device must be delivered sterile, sterilisation remains a separate step with its own process. We carry out disinfection and sterilisation in line with your requirements, from receipt of the parts to their final packaging.
Qualifying the room, then monitoring it: 2 separate controls
Qualification checks that a room reaches its class. Monitoring checks that it stays there between 2 qualifications.
Annex 1 asks that these 2 activities be clearly distinguished (section 4.24). Classification, which is part of qualification, counts particles of at least 0.5 µm and at least 5 µm, at rest and in operation, at a minimum number of sampling points given by ISO 14644-1 (sections 4.27 and 4.28).
Qualification also covers micro-organisms
Microbial contamination is expressed in colony forming units (CFU) and measured both at rest and in operation (section 4.31). For Grade D the annex sets these qualification limits (Table 2):
- 200 CFU per cubic metre of air sampled
- 100 CFU per 90 mm settle plate exposed for 4 hours
- 50 CFU per 55 mm contact plate
Personnel gowning serves to limit this contamination. In Grade D, Annex 1 asks for hair, beards and moustaches to be covered, and for a general protective suit and appropriately disinfected shoes or overshoes to be worn (section 7.13).
Requalification at least every 12 months
For Grades C and D the maximum interval between 2 requalifications is 12 months (section 4.32). Requalification includes, at a minimum, classification, the filter integrity test, airflow measurement and measurement of the pressure differential between rooms.
Between 2 requalifications, ISO 14644-2:2015 sets the minimum requirements of a monitoring plan for particle cleanliness. We entrust regular microbiological checks to an external laboratory.
7 points to check with a packaging provider
The classification displayed is not enough. Ask for the documents that prove it and those that show it is maintained.
- The latest classification report: date, measurement state (at rest, in operation), particle sizes, number of sampling points.
- The date of the last requalification, to compare with the maximum interval of 12 months in Annex 1.
- The monitoring plan and the limits adopted in operation, since Grade D sets none.
- The microbiological results: who samples, who analyses, how often.
- The flow of parts outside the room: cleaning, preparation, storage of work in progress and finished products.
- The separation of batches at the workstations.
- The link with traceability: each packaged batch must be traceable in a batch record.
Our batch record runs from raw material to shipment, with checks at receipt, during production and before release: see quality and traceability. All our services are presented on the medical device contract manufacturing page. The choice of the contract manufacturer itself is covered in our guide to ISO 13485 and contract manufacturing.
Sources for this guide
- ISO 14644-1:2015, Cleanrooms and associated controlled environments, Part 1: Classification of air cleanliness by particle concentration (overview page). International Organization for Standardization (ISO), accessed .
- ISO 14644-2:2015, Cleanrooms and associated controlled environments, Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration (overview page). International Organization for Standardization (ISO), accessed .
- EudraLex, Volume 4, EU Guidelines for Good Manufacturing Practice for Medicinal Products for Human and Veterinary Use, Annex 1: Manufacture of Sterile Medicinal Products, C(2022) 5938 final of 22 August 2022. European Commission, accessed .
- Good manufacturing practice guide, May 2024 (Annex 1, Manufacture of sterile medicinal products), in French. Agence nationale de sécurité du médicament et des produits de santé (ANSM), accessed .
- Publication and entry into force of Annex 1 of good manufacturing practice for sterile medicinal products, in French. Agence nationale de sécurité du médicament et des produits de santé (ANSM), accessed .
- Good preparation practices, 2023 (LD1, Preparation of sterile medicinal products, point 37 and note on classification), in French. Agence nationale de sécurité du médicament et des produits de santé (ANSM), accessed .
This article does not replace advice from a medical device regulatory professional.