Pad printing on medical devices: process, inks and checks

For manufacturers who have plastic parts or technical packaging marked for a medical device.

MarkingBy FrégatePublished 6 min read

Pad printing is an indirect printing process: a soft pad picks up ink from the engraving of a plate, then deposits it on the part, even a curved one. On a medical device the ink is one of the device’s materials: it enters its biological evaluation. The marking must adhere and stay legible throughout the intended lifetime, including after cleaning, disinfection or sterilisation cycles.

How does pad printing work?

The printing cycle breaks down into 4 steps:

  1. Ink fills the engraving of the plate, made of ceramic, photopolymer or steel.
  2. The excess is scraped off: ink stays only in the recesses.
  3. The pad presses onto the plate and picks up the pattern.
  4. It deforms against the part and deposits the ink.

The engraving depth is set according to the ink. The data sheet of a two-component Marabu ink recommends 20 to 24 µm. The same sheet allows a closed ink cup as well as an open one.

The pad conforms to the shape of the part. The process therefore marks small, curved surfaces that flat printing cannot cover.

What can pad printing mark on a medical device?

On small moulded parts, pad printing can mark, for example, alignment marks, part references, batch numbers, symbols or a logo. It also applies to technical packaging: tubes, cases, boxes.

Regulation (EU) 2017/745 on medical devices sets a principle that binds the device manufacturer: the information on the label must appear on the device itself. If that is not practicable or appropriate, all or part of it may appear on the packaging (Annex I, section 23.1 b).

The symbols used must comply with harmonised standards or common specifications (section 23.1 h). EN ISO 15223-1:2021, which specifies the symbols to be used with information supplied by the manufacturer, is on the list of standards harmonised in support of the Regulation, set by Implementing Decision (EU) 2021/1182. These symbols may be placed on the device itself, on its packaging or in the accompanying information.

At Frégate, pad printing follows moulding. We mould small precision parts, in engineering plastics including PEEK, then carry out trimming, finishing and pad printing. We also make and mark technical tubes, cases and boxes. See our page on medical injection moulding.

The ink is one of the device’s materials

The ink enters the biological evaluation and the choice of materials that the manufacturer justifies.

ISO 10993-1 frames the biological evaluation of medical devices within the ISO 14971 risk management process. It deals with how a device’s materials and components interact with biological tissue, taking into account their composition and the duration of exposure. Its 6th edition dates from November 2025 and replaces the 2018 one.

The Regulation lists 8 points that require particular attention (Annex I, section 10.1), 3 of which concern an ink directly:

  • the choice of materials and substances, particularly as regards toxicity
  • their compatibility with tissues, cells and body fluids, taking account of the intended purpose of the device
  • the effect of processes on the properties of the materials

The ink reference is therefore fixed in the marking specification, together with its hardener and thinner. Changing the ink amounts to changing a material of the device. It is a change to evaluate. The biological evaluation remains the device manufacturer’s.

Adhesion: prepare the surface, wait for the cure

Adhesion depends on the material, its preparation and the ink. Only tests on real parts demonstrate it.

Some materials can be pretreated. The Marabu data sheet mentioned above asks for polyethylene and polypropylene to be treated by flaming or corona discharge before printing. It states good adhesion from a surface tension of 42 to 48 mN/m and considers preliminary tests essential, because printability varies within a single type of material. On an engineering material such as PEEK, which we mould, only a series of tests on parts will show whether the chosen ink holds.

A two-component ink has a pot life

Two-component inks add a time constraint. For the ink cited, the mix with its hardener must be used within 16 hours, at 20 to 25 °C. Beyond that, adhesion and resistance may decline even if the ink still looks usable. The processing and drying temperature must not fall below 15 °C. Final hardness takes 5 to 7 days at 20 °C, or 30 minutes at 150 °C.

An adhesion test run on a print made the day before is therefore not conclusive.

The cross-cut test is one method among others

The test method is set in the specification. The cross-cut test of ISO 2409:2020 is one: a grid is cut through to the substrate, then the result is rated pass/fail or on a 6-class scale. ISO presents it as an empirical test, whose result depends on adhesion without measuring it. It is designed for paints and varnishes and is not suitable for coatings thicker than 250 µm. Its use on a pad printing ink is decided with the device manufacturer.

The marking must stay legible after every reprocessing

A reusable device goes through cleaning, disinfection or sterilisation cycles. The Regulation requires the carrier of its unique device identifier (UDI) to remain legible after each of them.

Annex VI, Part C, of Regulation (EU) 2017/745 requires reusable devices to carry their UDI carrier on the device itself. If they are cleaned, disinfected, sterilised or refurbished between 2 uses, this carrier must be permanent and remain legible after each operation, throughout the intended lifetime (section 4.10). The text provides 2 exceptions: direct marking that would compromise the safety or performance of the device, or marking that is technically impossible.

For any device, the medium, format, legibility and placement of the label must suit the device and its users (Annex I, section 23.1 a).

The durability of a marking is generally checked after the cycles planned for the device: its cleaning products, its disinfection protocol, its sterilisation process and the number of cycles over its lifetime. Legibility and adhesion are then assessed on the parts treated this way. We will agree with you at the project review which tests to run, at step 2 of our method: trials and validation.

Pad printing, laser or inkjet: how to choose?

Pad printing deposits ink by contact. The 2 other common processes follow a different logic.

Laser marking deposits nothing: it modifies the surface of the material, which requires a material that reacts to the beam. Inkjet projects ink without contact and can change the text from one part to the next, which suits variable data such as a serial number.

The choice depends on the shape of the area to mark, the material, the durability expected and the need for variable data.

A marking is documented like any manufacturing step

Marking is part of the batch. The materials and equipment used to make it, and the checks made on it, must be traceable.

Our batch record follows each batch from raw material to shipment, with checks at receipt, during production and before release. A certificate of conformity comes with every delivery. State in your specification what the record of a marking must show, for example:

  • the reference and batch of the ink, hardener and thinner
  • the plate used
  • the machine settings
  • the results of adhesion and legibility tests

Any change to these elements falls under the change control set out in the written agreement with your supplier. Our guide to ISO 13485 and contract manufacturing details what that agreement should cover.

What to gather before a marking request

A complete request comes down to a few documents and answers:

  • the part drawing, with the area to mark
  • the material and its exact reference
  • the pattern: texts, symbols, colour, size of the smallest characters
  • the planned contact with the patient or user and its duration
  • the cleaning, disinfection or sterilisation cycles the marking must withstand
  • the adhesion test method and its acceptance criterion
  • the planned volumes

With these elements we will advise you on the choice of materials, then move on to prototypes and tooling development before series production.

Sources for this guide

  1. Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC (Annex I, sections 10.1 and 23.1; Annex VI, Part C, sections 4.10 and 4.11). Publications Office of the European Union (EUR-Lex), accessed .
  2. Commission Implementing Decision (EU) 2021/1182 of 16 July 2021 on the harmonised standards for medical devices drawn up in support of Regulation (EU) 2017/745 of the European Parliament and of the Council (consolidated version of 17 June 2026). Publications Office of the European Union (EUR-Lex), accessed .
  3. ISO 10993-1:2025, Biological evaluation of medical devices, Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process (overview page). International Organization for Standardization (ISO), accessed .
  4. ISO 15223-1:2021, Medical devices, Symbols to be used with information to be supplied by the manufacturer, Part 1: General requirements (overview page). International Organization for Standardization (ISO), accessed .
  5. ISO 2409:2020, Paints and varnishes, Cross-cut test (overview page). International Organization for Standardization (ISO), accessed .
  6. Technical Data Sheet, Tampa Tech TPT, version 12, February 2020. Marabu GmbH & Co. KG, accessed .

This article does not replace advice from a medical device regulatory professional.